Instructions for Use

Acetabular Reamer Brushes

Reusable Nylon · Orthopedic Reamer Head Cleaning

SS-IFU-ACE-001 Rev. 10/2026
SurgiSafe™ by SurgiSite
Covered Family
Acetabular Reamer
Fill Material
Nylon
Sterilization
Steam · PreVac / Gravity
Use
Reusable · 10/Pack

1. Intended Use

SurgiSafe Acetabular Reamer Brushes are designed exclusively for cleaning acetabular bone reamers.

2. Quality Inspection

Ensure brush handles are not cracked, broken, or chipped before use. Discard and replace brushes once nylon bristles are worn or the brush has visible damage from repeated or prolonged use.

3. Storage

Store in a dry area where cross-contamination can be prevented.

4. Disposal

There are no special disposal instructions for this item beyond facility policy.

5. Known Issues

There are no known product issues or warnings.

6. Manual Cleaning

  1. Rinse brushes under cool water for one (1) to two (2) minutes (longer if needed) to loosen soil and remove gross debris.
  2. Prepare enzymatic solution according to the detergent manufacturer IFU. Soak brushes in fresh enzymatic solution for the manufacturer contact time (minimum two (2) minutes).
  3. Rinse brushes under cool water for one (1) minute.
  4. Visually inspect brushes for cleanliness and removal of all soil and gross debris.
  5. Repeat steps one (1) through four (4) if needed until brushes are completely free from visible soil.

7. Automated / Mechanical Cleaning

  1. Rinse brushes under cool water for one (1) to two (2) minutes (longer if needed) to loosen soil and remove gross debris.
  2. Prepare enzymatic solution according to the detergent manufacturer IFU. Soak brushes in fresh enzymatic solution for the manufacturer contact time (minimum two (2) minutes).
  3. Place brushes into a wire basket or micro-mesh tray, then place in an automated/mechanical washer.
  4. Visually inspect brushes for cleanliness and removal of all soil and gross debris.
  5. Repeat steps one (1) through four (4) if needed until brushes are completely free from visible soil.

8. Sterilization

Only new and unused brushes can be sterilized. Brushes may show slight discoloration from the sterilization process.

Cycle TypeCycle TemperatureMinimum Cycle Time
Pre Vacuum134°C (273°F)3 minutes
Pre Vacuum132°C (270°F)4 minutes
Immediate Use Pre Vacuum132°C (270°F)3 minutes
Gravity121°C (250°F)30 minutes

9. Additional Information

10. Covered Products

This IFU applies to all SurgiSafe Acetabular Reamer Brushes.

SS-ACE1

11. Contact

Questions on this IFU or survey documentation: process@surgisite.co · (770) 895-7050. Additional IFUs: SurgiSafe Instructions for Use.

SurgiSite, LLC McDonough, GA 30252
process@surgisite.co · (770) 895-7050
SS-IFU-ACE-001 · Rev. 10/2026