Instructions for Use

Channel Cleaning Brushes

Sterile Processing · Manual Cleaning of Lumened Instruments

SS-IFU-CCB-001 Rev. 10/2026
SurgiSafe™ by SurgiSite
Covered SKUs
All Channel Brushes
Shelf Life
5 Years from Manufacture
Use
Single Use Only
Standard
AAMI ST79 / ST91
Single Use Only — Do Not Reuse Reuse may result in cross-contamination, infection, or device failure. Discard immediately after use.

1. Description

SurgiSafe Channel Cleaning Brushes are made for manual cleaning of lumened surgical instruments and suction channels. Construction is twisted stainless-steel wire, nylon bristles, and a loop or ring handle. Diameters run from 1 mm to 24 mm, with lengths to match common instrument channel profiles. Packed 50/Pack unless otherwise labeled.

2. Intended Use

For manual pre-cleaning of reusable surgical instruments — lumens, channels, and suction tubes — before disinfection and sterilization. For use in sterile processing departments, GI labs, and ambulatory surgery centers in accordance with AAMI ST79.

3. Contraindications

4. Warnings and Precautions

Diameter Match An undersized brush may leave bioburden in the lumen. An oversized brush may damage the instrument channel or the bristles. Confirm diameter against the instrument manufacturer IFU before use.

5. Indications for Use

Use during the manual cleaning stage of the instrument reprocessing cycle to remove visible soil, tissue, blood, and debris from internal channels of surgical instruments, flexible endoscopes, suction cannulae, and lumened devices. Use as part of a validated reprocessing protocol in accordance with the instrument manufacturer IFU and applicable standards (AAMI ST79, AAMI ST91).

6. Storage Conditions

Store in a clean, dry environment at controlled room temperature (15°C–30°C / 59°F–86°F). Avoid excessive heat, humidity above 85% RH, direct sunlight, and chemical vapors. Keep in original packaging until point of use.

7. Shelf Life and Disposal

8. Directions for Use

  1. Confirm the brush diameter and length are appropriate for the instrument channel being cleaned. Consult the instrument manufacturer IFU if needed.
  2. Don appropriate PPE: fluid-resistant gown, gloves, and eye/face protection.
  3. Open the packaging in a clean environment. Inspect the brush. Discard immediately if bristles are damaged or packaging was compromised.
  4. Submerge the instrument and brush in a sink or basin of enzymatic detergent prepared per the detergent manufacturer instructions.
  5. Insert the brush gently into the lumen or channel. Using a back-and-forth motion, pass the brush the full length of the channel until the brush tip exits the distal end.
  6. Withdraw the brush and inspect the distal end for soil. Repeat until no visible soil is present on the brush tip or in the rinse.
  7. Rinse the instrument lumen thoroughly with clean water per facility protocol.
  8. After use, immediately discard the brush in a biohazard container. Do not set down, reuse, or attempt to clean, disinfect, or sterilize this brush.

9. Standards Alignment

10. Covered SKUs

This IFU applies to all SurgiSafe Channel Cleaning Brush SKUs, regardless of diameter, length, or pack. Current catalog SKUs:

SS-07039SS-16039SS-11055SS-11067 SS-09079SS-12079SS-16079SS-09091 SS-24098SS-09106SS-09118SS-12118 SS-16118SS-09130SS-09146SS-09157 SS-12157SS-16157SS-24157SS-08197 SS-12197SS-16197SS-24197SS-08236 SS-12236SS-16236SS-24236SS-08276 SS-16276SS-24276SS-08315SS-24315 SS-08393SS-12393SS-16393SS-24393 SS-08433SS-16433SS-08472SS-16472 SS-24472SS-16591SS-24591SS-08630 SS-08787SS-16787SS-24787SS-16886 SS-08944SS-16944

11. Contact

Questions on this IFU or survey documentation: process@surgisite.co · (770) 895-7050. Additional IFUs: SurgiSafe Instructions for Use.

SurgiSite, LLC McDonough, GA 30252
process@surgisite.co · (770) 895-7050
SS-IFU-CCB-001 · Rev. 10/2026